홈 >
액체크로마토그래피 사중극자 비행시간형 질량분석기(Ultra high speed-resolution Q-TOF LC-MS/MS System)
- 분석연구원
- 이상범
- 연락처
- 063-469-1904 / 010-4494-3446
- 설치장소
- 3305호
- 상태
- 신청가능
- 모델명
- X500R QTOF
- 제조사
- AB sciex
분석료
기본료 : 80,000원
• LC-MS(Positive, Negative) : 80,000원/점
(직접사용 25,000원/점)
• LC-MS/MS(Positive, Negative) : 100,000원/점
• peak search : 4,000원/peak
• 전처리(Vial, Filter 등) : 10,000원/점
(교내 50% 할인)
• LC-MS(Positive, Negative) : 80,000원/점
(직접사용 25,000원/점)
• LC-MS/MS(Positive, Negative) : 100,000원/점
• peak search : 4,000원/peak
• 전처리(Vial, Filter 등) : 10,000원/점
(교내 50% 할인)
세부분석/특기사항
(1) Mass Range : TOF 40 kDa or wider.
(2) Sensitivity, Positive MS/MS Quant Using the ESI probe in MS/MS mode :
reserpine 1 pg on column, S/N > 1,750:1 or better.
(3) Scan Speed, MS
Up to 25Hz (25 spectra per second) or faster
(4) Scan Speed, MS/MS
Up to 100 Hz (100 MS/MS spectra per second) or faster.
(5) Scan Speed, IDA
Up to 100 MS/MS spectra can be collected in 1sec or faster.
(6) Scan Speed, Sequential window acquisition of all theoretical mass spectra acquisition
Up to 200 MS/MS scans per cycle, in data-independent acquisition (DIA) or better.
(7) Resolution, MS High Mass : 42,000 (FWHM) at m/z 956 or better.
(8) Mass Accuracy, Internal : Less than 0.4ppm RMS or better
(9) Linear Dynamic Range : 4 orders or wider.
(10) Flow rate : 5ul/min to 2mL/min or wider.
(11) Automated calibration and tuning module.
(2) Sensitivity, Positive MS/MS Quant Using the ESI probe in MS/MS mode :
reserpine 1 pg on column, S/N > 1,750:1 or better.
(3) Scan Speed, MS
Up to 25Hz (25 spectra per second) or faster
(4) Scan Speed, MS/MS
Up to 100 Hz (100 MS/MS spectra per second) or faster.
(5) Scan Speed, IDA
Up to 100 MS/MS spectra can be collected in 1sec or faster.
(6) Scan Speed, Sequential window acquisition of all theoretical mass spectra acquisition
Up to 200 MS/MS scans per cycle, in data-independent acquisition (DIA) or better.
(7) Resolution, MS High Mass : 42,000 (FWHM) at m/z 956 or better.
(8) Mass Accuracy, Internal : Less than 0.4ppm RMS or better
(9) Linear Dynamic Range : 4 orders or wider.
(10) Flow rate : 5ul/min to 2mL/min or wider.
(11) Automated calibration and tuning module.